Error Correction 5 - 7
Siemens AG SPL5-330.840.01 Page 7 of 12 Uroskop ACCESS
Medical Solutions Rev. 01 03.03 CS SD 24 Endoscopy Option
Example of the test configuration with chrome-plated pliers, contrast OK.
Arrow 1:
The reflective areas of the pliers appear almost too bright and glaring and have a notice-
able white border, which indicates the beginning of camera overmodulation. The reflective
areas still indicate noise in the image.
Arrow 2:
When the monitor is set correctly, the structure of the pliers is visible without any satura-
tion effects.
Fig. 5