6.2.2 Installation of rotors and accessories
6.2.2.1 Installation of a rotor
Open the centrifuge lid by pressing the lid key.
Loosen the rotor tie-down screw by turning it anti-clockwise, but do not
remove it.
Lower the rotor with its central bore straight down onto the motor shaft.
Tighten the rotor tie-down screw clockwise with the supplied rotor
wrench with 3 Nm. In doing so, hold the rotor at its outer rim.
Follow the safety instructions and hazard warnings (see chapter 3 -
"Safety")!
WARNING
Once a day or after 20 cycles, the rotor tie-down screw must be loosened
by some turns, and the rotor must be lifted and fastened again. This
ensures a proper connection between the rotor and the motor shaft.
NOTE
Rotors can be used without a cover. This leads, however, to higher levels
of noise and temperatures when running.
Removal of a rotor
Loosen the rotor tie-down screw by turning it anti-clockwise and remove
the rotor.