“Cata / “Ana” modes:
All jewelleries must be removed.
Place the contact electrode covered with wet gauze, under the shoulder, when working
around face area or upper body or under the thighs when working on lower body.
Apply a layer of Sensibilis Lotion / Sensiblis Gel (post-depilation soothing gel for
“Cata” mode and pre-depilation lotion for “Ana” mode).
Use the roller electrode over the treated area for 3 minutes. (Never use metallic roller on
bare skin).
Remove gel and rinse the skin.
To prevent electrode deterioration, remove gel residues from roller electrode.
Use bactericidal wipes to disinfect roller.
Recommended intensity settings for Cata” mode:
For Face: 0.35 mA to 1.00 mA
For Body: 1.00 mA to 2.00 mA
" Treatment should be comfortable and soothing.
" Client should feel very light tingling or cooling sensation.
" A tingling may also be felt under the contact electrode (make sure the electrode is
covered with wet gauze).
Recommended intensity settings for “Ana” mode:
For Face: 0.30 mA to 1.00 mA
For Body: 1.00 mA to 2.00 mA
" Note: This phase should NEVER be used after electrolysis treatments.
" Client should only feel slight tingling. Lower intensity if client is uncomfortable.
" Treatment should never cause any discomfort.
" A tingling may also be felt under the contact electrode (make sure the electrode is
covered with wet gauze).