IV CALIBRATION ADJUSTMENTS
10
* Note’: When the instrument leaves the factory the password
is 123. However this can be changed as explained in Section
Vi:4.
In order to check to see whether a channel ha,s been adjusted,
proceed as follows:
Display Shows
,
IV: 2
1. Switch on
2. Press and hold ~
AUTO
3. Select channel, e.g. Number 1
4. Press ib/bu
kg/hi
(0.0 means that no adjustment has been
made).
HECTOLITRE (BUSHEL) WEIGHT
CAL.
IC
0.0
In the same way as moisture readings can be matched to oven
test results or some other test, e.g. merchants moisture
analyzer; the hectolitre (bushel) weight results can also be
adjusted to read the same as another instrument.
As for the moisture adjustment, it is important that the set-up
checks are correct, particularly the Weight Balance, before any
hectolitre (bushel) weight comparisons and any subsequent
adjustments are made.
The procedure to adjust the hectolitre (bushel) weight readings
is as follows:
Display Shows
1. Switch on