3
g) If the sample cell is not correctly mounted, the display will
show
Err.
after %H
2
0 is pressed.
If this appears on the display, remount sample cell firmly on
to the interface. Switch the Analyzer of f and then on again
before proceeding with your work.
h) Samples as high as 50°C. can be measured but do not leave the
instrument in direct hot sunlight. Clean the sample cell with
a soft brush or with a cloth dampened with methanol. NEVER USE
WATER! Take care not to damage the flexible temperature
sensor.
j) If you have more than one Analyzer, do not interchange sample
cells. Each Analyzer is individually calibrated with its own
sample cell for greater accuracy. Ensure that both the
instrument and the sample cell show the same serial number;
the serial numbers are located on the bottom of the
instrument and on the underside of the sample cell.
k) The SinarTM lM AP Moisture Analyzer has been calibrated using
clean samples. Accuracy can be affected by the presence of
stones, dirt, weeds, seeds, admixture, chaff or other foreign
matter.
111:2 MEASURING SAMPLE TEMPERATURE
Sample temperature can be measured by pressing TEMP when the
sample is in the measuring cell. If the sample temperature is
very different from the ambient temperature, leave the sample
for 30 seconds in the cell before you press the TEMP key.
Alternatively, keep on pressing the TEMP key until a stable
temperature is displayed.