IV CALIBRATION ADJUSTMENTS
11
4. Use the arrow keys to make the I 0.5
adjustment necessary for the
particular product,
e.g. if readings are to be increased by 0.5kg/hi,
press A 5 times.
5. Press CAL End I
AUTO
The adjustment is now complete.
In order to. see whether a channel has been adjusted,
proceed as follows:
Display Shows
1. Switch on I— HI — I
2. Select channel No. (1—7) e.g. i I
3. Press and hold P. CAL I
4. Press P I0.5
(Shows that readings have been increased by 0.5).
Note:
Please be aware of the effect of moisture content on
hectolitre (bushel) weight, i.e. above 15% as moisture
content increases the hectolitre (bushel) weight
decreases. For this reason it is best to base any
adjustment on tests which have been made using samples
which have a moisture content of approximately 15%.
V CALIBRATION TRANSFERS
The SinarTM T~~ AP Moisture Analyzer can be used to send
and receive calibrations from other SinarTM TM Moisture
Analyzers.
This facility is extremely useful for updating
calibrations and also for providing a calibration for
testing a new commodity.
The following instructions relate to communicating with
other AP Analyzers only. Details of communications with