13
Trial neck
Trial body
assembly
Distal reamer
Figure 9
PF Sleeveless Stem
Color A dimension
12/13 PF
17.5
14/15 PF
17.5
16/17 PF
19.5
18/19 PF
21.5
20/21 PF
23.5
22/23 PF
25.5
24/25 PF
25.5
26 PF
26.5
27 PF
27.5
The following information illustrates the approximate proximal size of each Sleeveless stem.
Trialing
Assemble trial components over the proximal
end of the distal reamer starting with the trial
body assembly and nally with neck trials
(Figure 9). At this point, establish correct femoral
anteversion. Check the trial body assembly for
soft tissue and/or bone debris that may impede
the components from seating properly.