Symptom Cause Remedy
No/intermittent
electrical power
Power cord not connected or
connected loosely at console
or electrical outlet
Note: The console will default
to power level 1 if power is
interrupted
Ensure that power cord is:
Fully seated into the back of the console and
electrical outlet
Not damaged and free of defects
For replacement power cord contact Customer
Care at 1-800-876-1261 (USA only) or local
Smith & Nephew representative
Power not present at
electrical outlet
Ensure that the electrical outlet has power
Connect to a different electrical outlet
Console power switch not
illuminated
Press power switch; power switch should
become illuminated
Footswitch LED
Footswitch not connected
properly
Observe red alignment dots on footswitch
connector and footswitch socket are properly
aligned
Ensure the footswitch connector is fully inserted
Console not
responding
to footswitch
and footswitch
connector indicator
is NOT illuminated
Footswitch obstructed Ensure that there are no objects obstructing the
footswitch from being depressed or releasing
Footswitch inoperative Order replacement footswitch (66800472) from
Customer C
are at 1-800-876-1261 (USA only) or
local Smith & Nephew representative
System fault LED
illuminated
Power fault caused
by over-current
1. Turn console OFF by pressing power switch
2. Ensure power cord is connected to an
appropriate power source. If necessary try a
different electrical outlet
3. Wait at least 5 seconds after turning OFF
4. Turn console ON by pressing power switch
System over-pressure
condition out of tolerance
Ensure the yellow high pressure tube on
handpiece is not kinked, obstructed or tangled
If error light is still illuminated, replace
handpiece
Contact Customer Care, 1-800-876-1261 (USA
only) or local Smith & Nephew representative to
arrange a return
Troubleshooting guide

Other manuals for smith&nephew VERSAJET II

Question and Answer IconNeed help?

Do you have a question about the smith&nephew VERSAJET II and is the answer not in the manual?

Summary

No;Intermittent Electrical Power Troubleshooting

Power Cord Connection Issues

Ensure the power cord is fully seated and undamaged. Contact support for replacements.

Electrical Outlet Power Check

Verify the electrical outlet has power and try connecting to a different outlet.

Console Power Switch Illumination

Press the power switch; it should illuminate when the console is on.

Footswitch LED Troubleshooting

Footswitch Connection Alignment

Observe alignment dots and ensure the footswitch connector is properly seated.

Footswitch Obstruction Check

Ensure no objects are obstructing the footswitch from being depressed or released.

Console Response Issues

Footswitch Inoperative Remedy

Order a replacement footswitch from Customer Care if it is not functioning.

System Fault LED Illuminated Troubleshooting

Over-Current Power Fault

Turn off, check power cord, wait 5 seconds, then turn console back on.

System Over-Pressure Condition

Check for kinks in the handpiece tube. Replace handpiece or contact support if error persists.

Handpiece Priming Issues

Fluid Supply Check

Ensure saline bag is full, flows freely, pinch clamp is open, and tube is clear.

Inflow Line Air Purge

Purge air from inflow tube by setting power to 10 and depressing footswitch.

Excessive Spray;Spattering Troubleshooting

Evacuation Orifice Obstruction

Remove any obstruction from the evacuation orifice (debris, tissue).

Waste Evacuation Tube Check

Ensure evacuation tube is connected, vented, not kinked, and container is positioned correctly.

Fluid Jet Striking Edge

Replace handpiece if fluid jet strikes the edge of the metal evacuation orifice.

smith&nephew VERSAJET II Specifications

General IconGeneral
Device TypeHydrosurgery System
Power Supply100-240 V AC, 50/60 Hz
Fluid ManagementIntegrated fluid management system
CategoryMedical Equipment
TechnologyHydrosurgery
Power SourceMains electricity
ApplicationWound debridement
HandpieceErgonomic, single-use
FunctionWound debridement
PortabilityPortable
TypeSurgical Debridement System

Related product manuals