Ensures personnel are trained, certified, and conversant with manuals and safety procedures.
Warns against oil, grease, and combustible lubricants near oxygen parts; regulators must not be immersed.
Verify paraPAC performance, including alarm functions, air entrainment, frequency, tidal volume, and O₂ concentration.
Detailed inspection of regulator assembly, including T-handle, yoke, filter, gauge, and relief valves.
Verify static and dynamic regulator output pressures using a test unit.
Check ventilator fittings, gauge, knobs, and connectors for security and smooth operation.
Verify correct alignment of Frequency, Tidal Volume, and Variable Relief knobs with calibration marks.
Inspect the patient valve assembly for cracks, splits, scoring, and valve element condition.
Test the main pneumatic switch for 'Off', 'Demand', and 'CMV/Demand' positions and alarm sequences.
Check the Air Entrainment NRV function by occluding the patient valve outlet.
Verify CMV/Demand function, breathing inhibition, and spontaneous breathing valve operation.
Check oxygen concentration of delivered gas with air mix using an oxygen monitor.
| Type | Portable Ventilator |
|---|---|
| Oxygen Concentration | 21 - 100% |
| PEEP | 0 - 20 cmH2O |
| Peak Flow Rate | Up to 120 L/min |
| Application | Emergency and Transport Ventilation |
| Ventilation Modes | CMV, CPAP |
| Patient Types | Adult, Pediatric |
| Power Supply | Internal battery |
| Battery Life | Up to 8 hours |
| Gas Supply | Oxygen, Air |
| Display | LCD |