20
Note: Clock Date and Time must be set before calibration interval will function. If batteries are removed
from wrench for longer than 20 minutes, clock will revert to default settings and must be re-
entered at power on.
Note: Calibration interval is calculated from either In-Service Date or last Calibration date (see SHOW
INFO menu) depending on which is more recent date. When clock Date is greater than In-Service
or Last Calibration date, plus Cal Interval, "CAL NEEDED" message will be displayed on power up
and after a re-zero. Pressing ENTER button will continue to target menu. Applying torque while
"CAL NEEDED" message is displayed will immediately display torque or angle measurement and
return to target menu when released.
Note: As an alternative to calibration interval, a Calibration Cycle Counter is provided in Calibration
menu (Contact your local Snap-on Repair Center regarding Calibration menu).
ENTER
button
CAL INTRVAL
DISABLED
EXIT
MODE SETUP
CALIBRATION
SET DATE/TIME

SET CAL INTRVAL
ENTER
button
UP/DOWN buttons
CAL INTRVAL
12 MON
EXIT