Grounding and Power Requirements
The TM5010/TM5030/TM5050 Series must be grounded. Improper connection of the
equipment’s grounding conductor can result in a risk of electric shock. Check with a qualied
electrician or service person if you are in doubt as to whether the outlet is properly grounded.
Do not modify the plug provided with the product or use a ground plug adaptor to adapt to a
non-grounded outlet. If the plug will not t the outlet, have the proper outlet installed by a
qualied electrician.
An extension cord should not be used between the equipment and the wall outlet.
The unit should not be plugged into a GFI plug or on a GFI circuit. The unit may not work or trip
the GFI circuit.
This product should only be run on the proper power designed for this unit. Check the power
label located on the front of the unit for proper power needs (see FIG. 1). To locate the power
outlet for your area please refer to FIG. 2. (example for the US and Canada, these units require a
115 V NEMA 5 - 15 non GFI circuit.)