User Manual SPA Snow Pack Analyser
- 10 -
The horizontal SPA-sensors are installed at predefined levels on the installation rack. The far end of the
sensors are connected at the far end of the framework. Then they are connected to the levers located on
the crossbeam of the mast. The stretching devices are fixed to the round cross-beam at the other end of
the framework. The springs are positioned directly at the levers and connected to the wires of the
stretching devices. Afterwards the sensors are spanned using the stretching devices until the spring has
elongated by about 100 mm and the sensor is 5000 mm long. The horizontal sensors are supported with
about 5 forks to ensure an upright and stable position.
3.7. Housing
Mount the housing on the mast and connect the cables of the SPA-sensors, the snow depth sensor and
the correction length sensor at the marked inputs.
3.8. Power supply
The supply voltage of the complete system is 10,5 to 15 VDC.
3.9. Summary
• Install framework with an initial tension.
• Locate snow depth sensor 1 m above the maximal expected snow depth.
• Install the sloping SPA-sensor higher than the maximal expected snow depth.
• Span the sensor using the stretching device.
• Install correction length sensor and set rotary bolt to correct position.
• Install and Span horizontal sensors by elongating the spring by about 100 mm.