User Manual SPA Snow Pack Analyser
- 16 -
Attention:
Approximately the first 7 minutes after connecting the SPA to the power
supply the measured values are not reliable if the SPA was disconnected for
more than a moment.
During this startup time invalid measurements are indicated with a value of
9999 pF for the measured low frequency capacity and 0000 pF for the
measured high frequency capacity.
example extened MIO protocol output:
I000000000000000000000409;
I0100-999-008-9990000043F;
I020099990000-9999999046B;
The 7 minutes is the time the internal high capacity electrolytic capacitors need to load. These capacitors deliver the
power for the measurement.