EasyManua.ls Logo

Somnetics Transcend - EN 60601-1 Compliance

Somnetics Transcend
59 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 50 Sleep Apnea Therapy User Manual
Appendix: Specifications
EN 60601-1 compliance
Protection against electric shock:
Degree of protection against ingress of water:
Use of flammable gasses:
Class II
Type BF
IPX0
Equipment not suitable for use in the presence of a flammable anesthetic
mixture with air or oxygen, or with nitrous oxide.
Interface physical characteristics
Interface (shell/seal) weight: 0.14 lbs (62 g)
Seal dead space Less than 301 cc
9” CPAP hose dimensions: 16-mm diameter, 20 cm long
6’ CPAP hose dimensions: 16-mm diameter, 1.9 m long
6’ Standard hose dimensions: 21-mm diameter, 1.8 m long
Performance
Sound power level (@ 10cm
H
2
O pressure, static)
37.0 dB
Sound pressure level (@
10cm H
2
O pressure, static)
29.0 dB
Resistance to flow of the
mask and application
accessories
0.140 hPa (cm H
2
O) @ 50 l/min
0.278 hPa (cm H
2
O) @ 100 l/min
Working pressure range:
4 to 20 cm H
2
O
Pressure limit:
30 cm H
2
O
Vent flow rate curve:

Table of Contents