A-1
Contents
Appendix A — Electromagnetic
Compatibility
Electromagnetic Emissions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Electromagnetic Immunity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Frequency Separation Distances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Electromagnetic Emissions
Note:
The 90217 ABP monitor has been tested under laboratory conditions and is suitable
for use in all establishments, including domestic establishments and those directly
connected to the public low-voltage power supply network that supplies buildings
used for domestic purposes. The customer, or user, of the module should ensure that
it is used in such an environment.
Electromagnetic Immunity
Note:
The ABP monitor is intended for use in the electromagnetic environment specified
below. The customer, or user, of the module should ensure that it is used in such an
environment.
Emission Test Compliance Electromagnetic Environment
RF emissions
CISPR 11
Group 1
Class B
The ABP monitor uses RF energy only for internal
function. Therefore, RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
6 kV contact
8 kV air
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.