Ambulatory Blood Pressure Monitor 90217 Service Manual
1-4
Office Check Mode
This mode verifies monitor operation and allows a user to view cuff pressure and blood pressure
results regardless of any previous disabling of the display. During Office Check Mode, the monitor
bleeds an additional pressure step below diastole.
This mode is enabled for the first five successful blood pressure readings (or attempted readings)
following monitor initialization. Office Check can be disabled by a patient cancel and can be re-
enabled after turning the power switch ON and holding the START/STOP button in while the last
digit of the software version number is displayed.
Blood Pressure/Heart Rate Measurements Displayed
These measurements can be selected to appear or not to appear at the end of a measurement.
Cuff Pressure Displayed
This measurement can be selected to appear or not to appear during the measurement cycle.