IM-P693-13 EMM Issue 2
7
CP42 Conductivity Probe
Install the probe as follows:
-
Ensure male and female threads are in good condition.
-
Use up to three turns (no more) of PTFE thread sealing tape on the probe thread.
WARNINGS: Do not use excessive tape. Do not use paste type jointing compound.
-
Fit and tighten the probe by hand initially. Use a suitable spanner to tighten the probe.
-
Due to the nature of a taper/parallel joint it is not possible to recommend tightening torque figures.
-
Do not overtighten - there should always be visible thread on the probe.
Note: The probe thread will not 'bottom out' (i.e. probe body hexagon contacts the face of the female
screwed connection), unless there is excessive wear or an out-of-tolerance female thread, in which case
it will be necessary to replace or re-work the flange or connection.
-
Take care not to damage the probe insulation sleeve when fitting the probe to the boiler.
Subsequent removal and refitting:
WARNING: Ensure boiler or vessel is depressurised and vented to atmosphere before attempting
to unscrew or remove the probe.
-
Always use the correct size spanner.
-
Inspect male and female threads for signs of damage, which may have occurred through overtightening,
leading to torn threads or even localised cold welding (galling/picking up).
-
If damage has occurred replace the probe.
-
Check that the probe insulation is not torn or damaged.