IM-P505-02-US Issue 1
50
UTM20 Series Ultrasonic Transit-time Flowmeters
10.5 Setup > Meter > Fluid
Fluid Submenus Options/Descriptions
FLUID
Water - Tap Acetone Ethylene Glycol
30%
Kerosene Propylene
Glycol 30%
Raw Sewage Ammonia Gasoline Methanol Stoddard
Solvent
Water -
Distilled
Benzene Glycerin Oil Diesel #1 Sulfuric Acid
96%
Water - Sea
3 5%
Ethanol Isopropanol Oil Diesel #2 Hydrochloric
Acid 36%
Brine - 3 5% Ethylene Glycol
100%
Jet Fuel A1/JP8 Propylene
Glycol 100%
Hyrdrofluoric
Acid 49%
Brine - 10% Ethylene Glycol
50%
Jet Fuel B/JP4 Propylene
Glycol 50%
Custom
CUSTOM FLUID
SOUND
SPEED
Numeric entry; Units ft/s or m/s based on velocity units
SPEED UNITS Ft/s or m/s
SPECIFIC
GRAVITY
Numeric entry; Specific gravity (density relative to water), pipe size
and viscosity are used to calculate the Reynolds number The Reynolds
number indicates whether the fluid is in turbulent, transition or laminar
flow and the flow profile
VISCOSITY
Numeric entry; Units centipoise (cP) or mPa-s Dynamic viscosity of
the fluid
VISCOSITY
UNITS
Units centipoise (cP) or mPa-s
REFERENCE
TEMP
Numeric entry, F or C Default 15 °C Reference temperature of viscosity
and specific gravity
REF TEMP
UNITS
F or C
SPECIFIC
HEAT
Numeric entry; Units: joule/gram °C; min 0 01, max 65 0; Specific heat
capacity is the heat capacity per unit mass of a material
10.6 Setup > Meter > Spacing
An asterisk (*) indicates the parameter default
Spacing Submenus Options/Descriptions
CALIBR ATED
SPACING
*Numeric display 0 to 300 units in" or millimeters, based PIPE settings
The spacing required between two transducers based on the pipe parameters Take
this measurement between the lines scribed into the side of the transducers or use
the scale on the rails, if used See the transducer user manual For Cx transducers
with fixed spacing, the parameter will not be shown