Warnings and Cautions alert users to dangerous con-
ditions that can occur if instructions in the manual are
not obeyed. Warnings are conditions that can cause
injury to the operator, while Cautions can cause dam-
age to the equipment.
WARNING:
Electrical shock hazard when connected to 300
V bias supply.
WARNING:
Exradin Ion Chambers are intended for quality
assurance and calibration measurements. They
are not intended to be connected to a patient for
real-time dosimetry measurements.
CAUTION:
Proper use of this device depends on careful
reading of all instructions and labels.
CAUTION:
Do not drop, mishandle, or disassemble unit
since it may result in a change of calibration
factor, or damage to the chamber. Refer all
servicing to qualied individuals.
CAUTION:
Do not sharply bend triax cable. Damage to the
cable may result in high leakage currents.
General Precautions