8
9
Collecting Volume:
Collecting Volume Length:
Collector Diameter:
Body Tube Outside Diameter:
Wall Thickness:
Chamber Length:
Body Tube and Guard Mate-
rial:
Response Uniformity over
the central 8.0 cm of chamber
length:
Collector Material:
Electrical Power Require-
ments:
Nominal Collection Efciency:
Maximum Polarizing Potential:
Nominal Inherent Leakage
Currents:
Low-Noise Triaxial Cable:
Connector:
Operating Parameters
Humidity:
Temperature:
Pressure:
Storage Parameters
Humidity:
Temperature:
Pressure:
Product Standards:
1.91 cc
8.0 cm
2.4 mm
12.7 mm
3.3 mm
17.0 cm
Shonka air-equivalent C552
plastic
±1.5%
Carbon Fiber
Operates at +300 VDC
100%
1000 VDC
10
-15
A
1.5 m long; 50 ohms; 29 pf/Ft
Triaxial BNC Plug (2-Lug
Male pin) and protective cap
connected by chain (standard);
others available upon request
20 to 80%, non-condensing
10 to 40
o
C
680 to 770 mm Hg
0 to 95%, non-condensing
-15 to 50
o
C
600 to 800 mm Hg
IEC 60601-1
1
, IEC 61674
1