10. After all numbers are installed, press the SET button and the displayed word “PRESET” will
again start to flash.
11. Finally, check to see that the numbers are correct and the slide is in the proper posi-
tion. Then press the PRESET button to install the whole number and the words “SET and
“PRESET” will go away.
Output
The height gage comes with an output port that allows data transmission to Starrett No. 772 Data
Collection Systems, No. 761 Multiplexer or to other devices that support Starrett RS232 format.
Cable No. PT62103 is needed to interface with the above Starrett devices.
Output is initiated via the button on the output cable or by transmitting an ASCII character to the
gage.
The output format is 4800 BAUD, 8 data bits, no parity, 1 stop bit, ASCII data. Transmission is
exactly 16 characters, followed by a carriage return and line feed.
Low Battery Indication
The display will show a dim reading when the batteries are getting weak. Replace with two new
batteries, CR2450 (No. PT61120).
Care and Maintenance
1. Do not store this instrument in areas of extreme temperature - below 10°F(-12°C) or above
130°F(54°C).
2. For best results, operate the height gage in an area free from dust and sprays of cutting oils
and other coolants.
3. This is a precision instrument and extreme care should be taken to avoid bumping or drop-
ping the gage.
4. The height gage has been calibrated at 68°F(20°C) (see specifications). Avoid using the
gage in direct sunlight or in a draft of hot or cold air. For best results, use in a constant temp
perature situation and allow time for the gage to soak at temperature for at least one hour
when transferring the tool from a temperature difference of more that 10°F(-12°C).
5. Clean the height gage vertical bar using a clean, lint-free cloth. Apply a very light coat of
lubricant. Use one or two drops of instrument oil on a finger and wipe over the vertical bar.
Move slide up and down and wipe off the excess oil.
SPECIFICATIONS
I
NCH MILLIMETER
Measuring Range 12” model 0-12 0-300
24” model 0-24 0-600
Accuracy 12” model ± .0015 ± 0.04 mm
24” model ± .0015 (0-18”) ± 0.04 mm
± .002 (18”-24”) ± 0.05 mm
Resolution .0005” .01 mm
Power Source Batteries - Lithium Coin Cells 3V
(2) No. CR2450 (No. PT61120) or Equivalent
1-1/2 Years Battery Life with Average Use