MATTRESS CPR VALVE CHECK
After approximately
30-40
minutes
the
mattress should
be
fully inflated.
If the
mattress
is NOT
fully
inflated,
check
that the CPR valve is in the
closed position
and that the
pump
is
connected
to the
mattress
correctly.
"
"
"
"
CAUTION: Please ensure that the PU mattress cover is fully zipped and CPR connection
is
closed properly after
inspection.
CAUTION: During a power cut, pump will stop functioning and the ‘power failure LED’ indicator
may be flashing, and the power failure alarm may be triggered if equipped; this is normal. The
pump will return to its normal operation when power is resumed.
"
MATTRESS COMFORT CHECK
Re-check the
comfort control setting
on the
pump
unit.
If
required, adjust
the
control setting
to
increase
or decrease the
required rmness
of the
mattress,
using
the
patients weight
as an
initial
guideline.
If the
patient
back rest
moves greater
than 30 degrees,
increase
the
pressure
in the
mattress
to
accommodate
the extra load on the
sacral
cells.
HAND CHECK
To
check whether
the
patient
is
being properly supported, slide
one hand in between the air
mattress
and
the base to
ensure
that they are
not ‘bottoming
out’.
A gap of 2-4 cm /1-2
inches
between the
patients bottom
and the bed base is an
acceptable
gap and is an
indication
of the
correct pressure setting.
If
this
is
not
the case, the
pressure
can be
adjusted
accordingly,
using
the
comfort control
dial.