EasyManua.ls Logo

Steiss UTS - Specification; Technical Specifications and Accessories

Default Icon
24 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9. Specication
1111
Model name: Steiss™ Universal Therapy System (UTS)
Size in CM (L x W x H):
200 x 90 x 18 cm
Weight of system: 15kg
Cycle time (min): 10 min
Max user weight: 175kg
Minimum user weight: 30kg
Min/Max pressure: 20 ~ 60 mmHg +/- 6mmHg
Max ow rate:
6 L/min
Rated voltage: AC100 – 240V / 50Hz-60Hz
Max current: 0.2-0.1A
Fuse rating: T1AH 250V
Protection type: Class II Type BF
Ingress of water protection: IP22
Mode of operation: Continuous
Environment (Temperature):
Operation: 15ºC to 35ºC (59ºF to 95ºF)
Storage: 5ºC to 60ºC (41ºF to 140ºF)
Environment (Humididty):
Operation: 30% to 75% non-condensing
Storage: 30% to 90% non-condensing
Operation atmospheric pressure range:
700 hPa to 1060 hPa
Operation altitude:
-1017 feet to 9,843 feet (-310 metres to 3000 metres)
Test standard: IEC60601-1, IEC60601-1-2 and IEC60601-1-11
Safety standard: CE
Mattress material:
Top Cover: Two way stretch, PU laminated Nylon
Base Cover: PVC laminated Polyester
Packaging in CM (L x W x H): 67 x 27 x 98 CM
Item Model Material
Handset HS-NBE ABS
Accessories: