7 Troubleshooting
66
Operator Manual Amsco® 3085 SP Surgical Table
150830353_AB
7.15 Surgical Table Does Not Articulate nonALM
Condition
Operator cannot articulate Surgical Table.
General Troubleshooting
Additional Information
Table unplugged (electric-powered Table only).
1. Plug in Table.
No Facility power (electric-powered Table only).
1. Have Facility turn power on.
Power cord defective (electric-powered Table only).
1. Replace power cord.
Batteries totally discharged (battery-powered Table
only).
1. Recharge batteries.
Circuit breaker CB-1 tripped (electric-powered Table
only).
1. Reset CB-1 circuit breaker.
Circuit breaker CB-2 tripped (electric-powered Table
only).
1. Reset CB-2 circuit breaker.
7.16 Surgical Table Motor Batteries Do Not Charge nonALM
Condition
User attempts to charge Surgical Table batteries and batteries do not charge.
NOTE: This alarm is for battery powered Surgical Tables only.
General Troubleshooting
Additional Information
Power cord defective.
1. Replace power cord.
Circuit breaker(s) tripped.
1. Circuit breaker CB-4 tripped reset.
2. Circuit breaker CB-2 tripped reset.
3. Contact STERIS.
Electrical System malfunction.
1. Contact STERIS.
7.17 Table Lash Appears to be Excessive nonALM
Condition
Table lash appears to be excessive.
General Troubleshooting
Additional Information
Lash requires inspection.
1. Contact STERIS.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Steris Amsco 3085 SP and is the answer not in the manual?

Steris Amsco 3085 SP Specifications

General IconGeneral
BrandSteris
ModelAmsco 3085 SP
CategoryMedical Equipment
LanguageEnglish

Related product manuals