Note
Make sure that the battery pack is fully charged prior to use: the LED indicator is
green.
Note
The battery pack does not need to be inserted during charging.
5.3 Power supply
ELECTRICAL HAZARD
The machine must be earthed (grounded).
Make sure that the actual electrical power supply voltage corresponds to the voltage
stated on the type plate of the battery pack.
Incorrect voltage can damage the electrical circuit.
Note
The equipment is shipped with 2 types of electrical power cables. If the plug supplied
on these cables is not approved in your country, the plug must be replaced with an
approved plug.
The European Schuko plug
The 2-pin plug (European Schuko) is for use on single-phase
electrical power connections.
The leads must be connected as follows:
Yellow/Green Earth (ground)
Brown Line (live)
Blue Neutral
The North American NEMA 5-15P plug
The 2-pin plug (North American NEMA 5-15P) is for use on
single-phase electrical power connections.
The leads must be connected as follows:
Green Earth (ground)
Black Line (live)