Stryker 941 - Technical Specifications

Stryker 941
21 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12
Model REF 940
Electrical: 100-120V~ 50-60Hz 1.5A
Speed: Nominal at 100V~
Nominal at
120V~
High 12,800 cpm*
High 15,900
cpm
Low 7,700 cpm
Low 10,700
cpm
DOUBLE INSULATED
Class II equipment
Type BF Applied Part
IPX0 Ordinary Equipment
Duty Cycle:
Time on–15 minutes
Time off–20 minutes
Weight: 2.8 lbs. (1.3 Kg)
Length: 10.0 in. (254 mm)
Height: 3.0 in (76 mm)
NOTE: Only the items listed in the Replacement Parts List are
user serviceable. In case of other operating difficulties, return to
Stryker for repair.
CSA International
• CAN/CSA-C22.2 No. 601.1 M-90
• IEC 60601-1
• UL 60601-1
Model REF 941
Electrical: 220-240V~ 50-60Hz 0.75A
Speed: Nominal at 230V~ 50Hz
High 15,900 cpm*
Low 10,700 cpm
DOUBLE INSULATED
Class II equipment
Type BF Applied Part
IPX0 Ordinary Equipment
Duty Cycle: Continuous Operation with intermittent
loading.
Time on–15 minutes
Time off–20 minutes
Weight: 2.8 lbs. (1.3 Kg)
Length: 10.0 in. (254 mm)
Height: 3.0 in (76 mm)
NOTE: Only the items listed in the Replacement Parts List are
user serviceable. In case of other operating difficulties, return to
Stryker for repair.
Meets applicable requirements of:
IEC 60601-1 Safety Requirements
* Cycles per minute.
Specifications
If the equipment experiences sporadic electrical interference:
Turn off all electrical equipment not in use.
Relocate electrical equipment; increase spacial distance.
Plug equipment into different outlets.