There are abnormal
pressure or ow rate
uctuations�
There is a pressure sensor error Reboot the system�
If the problem persists, contact your
Stryker representative, or return the
console for repair
A hardware fault is detected�
The pump stops
pumping uid, and
the pressure indicator
continues to blink�
The pressure is too high� The actual
pressure exceeds 200mmHg for
15seconds or greater than 250mmHg
for ve seconds�
Open the valve at the outow tube, the
drainage tube, or the stopcock on the
instrument to reduce the pressure�
The pinch valves are
not engaging�
There is a pinch valve or hardware
error
Examine the pinch valves to ensure
they are functioning properly
Reboot the system�
Bubbles appear in the
joint�
There is no more irrigation uid� Replace or add additional saline bags�
Ensure the pinch clamp is open on the
irrigation tube, or the stopcock is open�
The tubing connection is loose� Ensure the tubing is securly connected�
If the problem persists, replace the
tubing�
A priming error occurred� Ensure the pinch clamps on the saline
bags are open�
Ensure the Inow or Outow Cassette is
fully inserted�
Remove the hardware from the tubing
and repeat the priming step
If the problem persists, replace the
cassette�
The suction level is set too high� Decrease the suction level�
The pump cannot
achieve the set
pressure
There is a hardware setup error, the
ow limit is set too low, or the suction
level is set too high�
Verify the luer-lock is tightly closed,
the correct hardware is selected and
properly connected, and the dual
stopcock cannula is properly set up
Increase the ow limit or the set
pressure setting
Decrease the suction level�
Press Run/Stop to restart the pump�
There is insucient
pressure in the surgical
site�
An irrigation problem exists� Check the stopcock on the arthroscope
and the clamps under the saline bags�
The hardware set up is incorrect� Ensure the hardware is properly
selected and set up�
There is no suction
while the pump is
running�
The approved resection consoles are
improperly connected�
Ensure the consoles and the tubes
are properly connected according to
the instructions in Setup and Device
Conguration section of this manual�
The tubes on the Outow Cassette are
improperly connected�
Unable to upload les
to/from USB drive
A hardware fault occurred� Ensure the USB drive is functioning
properly and the correct les are on
the drive� If the drive is not functioning
properly, replace the USB drive
Reboot the system�

Other manuals for Stryker CrossFlow

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker CrossFlow and is the answer not in the manual?

Stryker CrossFlow Specifications

General IconGeneral
BrandStryker
ModelCrossFlow
CategoryWater Pump
LanguageEnglish

Related product manuals