EasyManua.ls Logo

Stryker CrossFlow - Interference, Touchscreen, and File Issues; Stuck Cassette and Service Notifications

Stryker CrossFlow
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EN-31
DRAFT
Sporadic electrical
interference is aecting
the pump�
Electrical, RF, and/or mobile
communications equipment is
aecting the normal function of the
pump�
Power down all electrical equipment
not in use�
Increase the distance of other electrical
equipment�
Connect the pump and other
equipment into dierent outlets�
The touch screen
is unresponsive or
inaccurate
Touch screen is not properly calibrated
for the user
Reboot the system�
If the problem persists, contact your
Stryker representative or return the
console for repair
A hardware or software error occurred�
A user preference
le was accidentally
deleted from the pump
Unintentional deletion of a user
preference le�
Transfer the user preference le from
a USB drive to the pump according to
the instructions in the Transfer Files
to/from a USB drive section of this
manual�
The pump is stopped,
but the Outow
Cassette is stuck and
cannot be ejected�
The pinch valves do not retract� Do not attempt to remove the cassette
as it may damage the pump or the
cassette�
Reboot the system�
If the problem persists, contact your
Stryker representative
“Service Pump Soon The pump is 95% through the current
maintenance period�
Contact your Stryker service
representative�

Other manuals for Stryker CrossFlow

Related product manuals