Sporadic electrical
interference is aecting
the pump�
Electrical, RF, and/or mobile
communications equipment is
aecting the normal function of the
pump�
Power down all electrical equipment
not in use�
Increase the distance of other electrical
equipment�
Connect the pump and other
equipment into dierent outlets�
The touch screen
is unresponsive or
inaccurate
Touch screen is not properly calibrated
for the user
Reboot the system�
If the problem persists, contact your
Stryker representative or return the
console for repair
A hardware or software error occurred�
A user preference
le was accidentally
deleted from the pump
Unintentional deletion of a user
preference le�
Transfer the user preference le from
a USB drive to the pump according to
the instructions in the Transfer Files
to/from a USB drive section of this
manual�
The pump is stopped,
but the Outow
Cassette is stuck and
cannot be ejected�
The pinch valves do not retract� Do not attempt to remove the cassette
as it may damage the pump or the
cassette�
Reboot the system�
If the problem persists, contact your
Stryker representative
“Service Pump Soon The pump is 95% through the current
maintenance period�
Contact your Stryker service
representative�

Other manuals for Stryker CrossFlow

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker CrossFlow and is the answer not in the manual?

Stryker CrossFlow Specifications

General IconGeneral
BrandStryker
ModelCrossFlow
CategoryWater Pump
LanguageEnglish

Related product manuals