EENN
www.stryker.com 2971-109-0 01 REV B 1-93
EMC Information
(Continued)
Guidance and Manufacturer’s declaration - Electromagnetic Emissions
Isolibrium is intended for use in an electromagnetic environment specified below. The customer or the user of
Isolibrium should make sure that it is used in such an environment.
Emissions Test
Compliance Electromagnetic Environment
RF Emissions
CISPR 11
Group 1
Isolibrium uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF Emissions
CISPR 11
Class A
Isolibrium is suitable for use in all establishments
other than domestic and those directly connected to
the public low voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations
Flicker Emissions
IEC 61000-3-3
Complies
WARNING
This equipment/system is intended for use by healthcare professionals only. This equipment/system may cause radio
interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures, such as
reorienting or relocating Isolibrium or shielding the location.