Inspection checklist
Refer to the Operating
Instructions for the
User’s Checklist
Inspect the debrillator for
signs of damage, cracks or
foreign substances
It is not necessary to switch
on the debrillator for the
inspection
Observe the Readiness Display
on the top right and perform
the following checks:
1. The OK symbol is visible
2. The battery symbol shows
two or more bars
Conrm spare, fully charged
battery is readily available
and in-date
Ensure there is one pack
of QUIK-COMBO pads
preconnected to the
debrillator, and the pack of
pads is sealed and in-date
There should be one
additional set of adult pads
If present, check Infant/Child
Reduced Energy Debrillation
Electrodes to make sure they
are sealed and in-date
This is not the complete instructions for use. Refer to
Operating Instructions for complete device information
including directions for use, intended use and warnings.
Readiness
display
Battery is very low.
Replace immediately.
Battery is depleted.
Replace immediately.
Service symbol indicates
service is required.
Battery is fully charged.
OK indicates device
is ready for use.
LIFEPAK
®
1000 Defibrillator
Pocket guide

Other manuals for Stryker LIFEPAK 1000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker LIFEPAK 1000 and is the answer not in the manual?

Stryker LIFEPAK 1000 Specifications

General IconGeneral
BrandStryker
ModelLIFEPAK 1000
CategoryMedical Equipment
LanguageEnglish

Related product manuals