EasyManuals Logo

Stryker Neptune 3 User Manual

Stryker Neptune 3
63 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #54 background imageLoading...
Page #54 background image
0703-001-700 Rev-C EN
www.stryker.com 53
CODE MESSAGE PRIORITY ACTION
12.X Level
Sensor
Error
MEDIUM Call service.
13.X Vacuum
Error - 20L
MEDIUM Call service.
13.6 20L
Canister
Pressure
LOW Turn on suction to
reduce the positive
pressure in the
canister. If this
notification persists,
call service.
14.X Vacuum
Error - 20L
MEDIUM Call service.
15.X Vacuum
Error - 4L
MEDIUM Call service.
15.6 4L Canister
Pressure
LOW Turn on suction to
reduce the positive
pressure in the
canister. If this
notification persists,
call service.
16.X Vacuum
Error - 4L
MEDIUM Call service.
18.3 4L Almost
Full
LOW Prepare to use an
alternate source of
suction, or empty the
4L canister of fluid
waste to prevent the
loss of suction.
18.4 20L Almost
Full
LOW Prepare to use an
alternate source of
suction. 20L suction
will stop when the 20L
canister is full.
18.5 4L Full MEDIUM Use an alternate
source of suction, or
empty the 4L canister
to restore lost suction
to the 4L canister.
18.6 20L Full MEDIUM Use an alternate
source of suction,
or dock the rover
to empty the 20L
canister.
CODE MESSAGE PRIORITY ACTION
20.1 Audio
Failure
MEDIUM Cycle power to the
rover. Push the power
switch to the OFF
position; then push
the switch to the ON
position. If this error
persists, call service.
21. 0 Knob/Dial
Error
MEDIUM Call service.
22.0 4L
Manifold
Error
MEDIUM Call service.
22.1
22.2
23.0 20L
Manifold
Error
MEDIUM Call service.
23.1
23.2
24.11 4L
Manifold
Error
LOW Install new manifold.
24.13 Expired 4L
Manifold
LOW Install new manifold.
24.15 Used 4L
Manifold
LOW Install new 4L manifold
before performing
the next surgical
procedure.
25.11 20L
Manifold
Error
LOW Install new manifold.
25.13 Expired
20L
Manifold
LOW Install new manifold.
25.15 Used 20L
Manifold
LOW Install new 20L
manifold before
performing the next
surgical procedure.
0000160552, Rev. C Effective Date: Nov 25, 2015 1:43:15 PM
Print Date: Nov 25, 2015 02:19:13 PM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Neptune 3 and is the answer not in the manual?

Stryker Neptune 3 Specifications

General IconGeneral
BrandStryker
ModelNeptune 3
CategoryWater Pump
LanguageEnglish

Related product manuals