6390 Angle Position Sensor (APS) Calibration Instructions
TOOLS REQUIRED
• StrykerEMSUSBTool • Male-to-MaleUSBCable
• T25TorxDriver • TapeMeasure
PROCEDURE
1. Placethevehicleonalevelsurface.
2. MakesurethatthePower-LOADmainpoweristurnedon.
3. PressandholdthereleaseleveratthefootendofthePower-LOADsystemandpulltoremovethecotfromthe
vehiclepatientcompartment.
Note: The head end LED indicators turn solid green only when the cot is ready to unload.
4. Unloadthecotfromthevehicle:
• For cot models 6500/6506 & 6510/6516:
1. Pressandholdtheextend(+)buttononthecotcontrolswitchtoextendthecotbaseuntilthecotwheels
restontheground.
2. Releasetheextend(+)buttonafterthecotisnolongersupportedbytheliftingarms.Theliftingarmswill
continuemovinguntiltheyhavefullylowered.Do not unlock the cot from Power-LOAD.
• For cot models 6085/6086:
1. Operator1(footend)-Graspthecotframeatthefootend.Squeezeandholdthecotmanualrelease.
2. Operator2(side)-Grasp the baseframe,liftslightly,andlowerthebase frame toits fullyextended
position.Verifythatthebaseisfullyextended.
3. Operator1(FootEnd)−Releasethe cotmanualreleasetolock theundercarriageintotheextended
position.
4. Pressthedown()buttononthePower-LOADcontrolpaneltolowertheliftingarmsandcot.Do not
unlock the cot from Power-LOAD.
5. UsingaT25Torxdriver,removethetwoscrews(A)thatsecurethecovertotheleftwingassemblytogainaccess
to the USB port (Figure 1). Save the screws and
coverforreinstallation.
6. UnscrewthecapthatprotectstheUSBport.
7. Set the laptop with the Stryker EMS USB tool
installedontopofthecotmattressatthefootend
ofthecot.
A
Figure 1