Cot gets stuck
when unloading
Scenario
Cot gets stuck halfway out when you
press the red release lever at the foot
end of the anchor.
1. Pull the patient left transfer lock
trigger and push the cot back into the
vehicle until it locks into place.
2. Hold down the red release lever at the foot
end of the anchor with your thumb and use
your knuckle to hold the transfer in place
while you pull the cot toward you until it
raises up. Pull the cot out to unload.
Transfer lock
triggers
Transfer assembly
Red release
lever
Trolley assembly