700000544314 Rev-AD EN
www.stryker.com 11
Power Cord (general)
Ratings:
Current: 10 A
Voltage (nominal): 250 VAC minimum
Frequency: 50-60 Hz
Temperature: 0 °C to 70 °C minimum
Flammability: UL 94 V-2 minimum, IEC 60332-1
Conductor:
Size: 3 X 1.00 mm
2
≤ size < 3 X 1.50 mm
2
Material: Copper
Connector Type: IEC 60320 C13
Plug Type: The facility power (mains) plug shall have a ground/earthing pin.
Cord:
Type: SJT, H05VV-F, HVCTF, RVV or equivalent (unshielded)
Length: 3.0 m, 2.5 m
NOTE – The 2.5 m cord is not for use in Canada or the US.
Dielectric Withstand: 1500 VAC for 60 seconds between line and protective earth, and between neutral and protective earth
Certification: All applicable in-country medical electrical requirements
Power Cord (country-specific)
The Canadian and US power supply cord shall have a tag or label in English and French indicating that “GROUNDING RELIABILITY CAN ONLY BE ACHIEVED WHEN
EQUIPMENT IS CONNECTED TO AN EQUIVALENT RECEPTACLE MARKED ‘HOSPITAL ONLY’ OR ‘HOSPITAL GRADE’ ” or equivalent wording.
Agency Approval: CSA Certified for Canada and US or UL Recognized for Canada and US.
Environmental Limitations Operation Storage and Transportation
Temperature:
Humidity:
Atmospheric Pressure: