18. Disconnect the charge line from the adapter fitting. Store the charge line to keep it clean.
19. Remove both adapter fittings.
NOTE
Reconnect the adsorber if it was disconnected prior to gas cleanup.
20. If other components need cleaning, refer to the procedures in their technical manuals.
Otherwise, reconnect the supply and return gas lines to the compressor. Torque the gas
line couplings to 4.85 0.7 kgf m (35 ± 5 ft. lbs.).
21. Leak check the Aeroquip couplings. See the Leak Check procedure.
This completes the gas cleanup procedure for the compressor.
Helium Leak Check
In addition to identifying suspected leaks, check the compressor for helium leaks each time
it has undergone any amount of disassembly. Use a helium mass spectrometer leak detector if
available. Follow its manufacturer’s instructions.
If a leak detector is not available, use a commercial leak detection solution. However, small
leaks may not be detected. Also, it is important to:
fully coat the joint being tested;
allow time for bubbles to form at a small leak;
look carefully for the smallest bubble formations.
After solution testing is completed, use water to wash all residues from joints and couplings.
In either method of testing, do not assume that one leak is the only one. Check all joints.
Leak detection by instruments can be misleading. Leaking gas can form patterns that indicate
leaks at sound joints. Large leaks or a high gas concentration can make isolation difficult.
Leaks occur most frequently at threaded joints. However, they can occur also at brazed and
welded joints.
The flat gasket in the face of the Aeroquip male coupling seals the joint. A leak at this gasket
seal can be detected only when a gas line is connected. A leak here can be caused by:
the coupling not fully tightened;
a worn, damaged, or missing gasket seal;
dirt on or under the gasket seal;
dirt on the female coupling’s mating surface;
damaged parts on either coupling which prevent proper mating or sealing.
Helium Leak Repair
Tools required: Open end wrenches 1", 1 1/8", 1 3/16"
#2 Phillips screwdriver
Leaks in flexible metal tubing cannot be repaired. Discard the damaged gas line and install
a new one.