EasyManua.ls Logo

Supelco Spectroquant Prove Series - Getting Started and Initial Setup; Initial Setup Procedure

Supelco Spectroquant Prove Series
26 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Version 2.0 – 08/2018 11
4
5
2
1
3
3 Getting Started
3.1 General Notes on Handling
The Spectroquant
®
Prove spectrophotometer is
an optical precision instrument. It should always
be handled with care, especially when in mobile
use. Always protect the instrument from condi-
tions that could damage the mechanical, optical
and electrical components. Please note the fol-
lowing in particular:
The temperature and humidity during opera-
tion and storage must be within the limits
specified in the "Technical Data" section (see
the Operating Manual)
The instrument must never be exposed to
the following:
Extreme dust, humidity and moisture
Intense light and heat
Fumes that are corrosive or contain high con-
centrations of solvents
In addition take care of the following:
For measuring, the instrument must be placed
on a flat surface
Spilled liquid or other material should be re-
moved immediately (see the Operating
Manual)
If a cell has broken in the cell holder, the cell
holder should be cleaned immediately (see the
Operating Manual)
The cover should always be closed when the
spectrophoto meter is not in use
When the spectrophotometer is being trans-
ported, the cell compartment must be empty
3.2 Initial Setup
Proceed as follows:
Connect the power adapter (see section 3.2.1)
Switch on the spectrophotometer (see section
3.2.2)
Set the language (see section 3.2.3)
Set the date and time (see section 3.2.4)
Run the self-test (see section 3.2.5)
NOTE
For our Operating Manual please visit:
www.sigmaaldrich.com/spectroquant
For more information about the technical videos
please visit:
www.sigmaaldrich.com/photometry
3.2.1 Connecting the Power Supply
Power is supplied through the power adapter
provided. The power adapter supplies the spec-
trophotometer with the required voltage and
type of current (24 V DC).
CAUTION
The line voltage at the user location must fulfil
the specifications stated on the power adapter
(the specifications are also indicated in the Op-
erating Manual). Only ever use the 24 V power
adapter provided. Please note that damage
caused by using a different power adapter than
the one supplied voids all warranty claims.