4 Operation – 4.4 Measurements
4.4.1 Performing a Measurement
Measurements can be performed using rect-
angular cells of various path lengths (10, 20,
50 mm/100 mm Prove 600) and Spectroquant
®
round cells. Insert cells as follows to start the
measurement:
Measuring with a round cell with closed lid
Measuring with rectangular cells with open
lid:
insert AutoSelector
Open the flip-up cover
4
by pushing it back
with your fingers
Insert the AutoSelector vertically into the cell
compartment
5
, ensuring that the white posi-
tion mark
6
on the AutoSelector is aligned with
the positioning mark on the spectrophotometer
7
The photometer is ready to measure
NOTE
If the barcode cannot be read, please see the
Operating Manual.
Insert the barcoded Spectroquant
®
round cell
through the opening
1
, ensuring that the
white position mark
2
on the cell is aligned
with the positioning mark on the spectrophoto-
meter
3
Measurement starts automatically, and the
measurement result is displayed in the con-
centration measurement overview (see page
7)