E
21
The other commands are shown in the following table:
AOFF 0 / AOFF 1 Deactivates/activates the automatic switch-off
AOFF? Forwards the state of the automatic switch-off
BAT? Sends the status of the battery (BAT1=ok or BAT0=low)
ID ? Sends the instrument identification (SY276)
IN Activates the Inch unit
KEY? Sends keyboard status (KEY0 or KEY1)
KEY 0 / KEY 1 Locks/unlocks the keyboard (except data output)
ECO 1 / ECO 0 Activate/de-activate economic mode
ECO? Current economic mode
LCAL? Sends the date of the last calibration (DD.MM.YYYY)
LCAL DD.MM.[YY]YY Sets the date of the last calibration
MM Activates the millimeter unit
16. Remote commands
The data request is done by the sending of the character ASCII <?>, followed by <CR> (carriage return). The instru-
ment also answers in the same way during the detection of an impulse pedals/automat.