MAINTENANCE AND SUPPLIES REPLACEMENT
4-14 Sysmex KX-21 Operator’s Manual -- Revised October 1998
(4) Using the filler provided with the unit, pour approximately 1 mL each of
CELLCLEAN into the WBC transducer and RBC transducer.
Filler
Figure 4-5-8: Pouring CELLCLEAN
Since CELLCLEAN is a strong alkaline detergent, take care not to have
it adhere to the skin or clothes. If the skin or clothes should come in
touch with it, flush it away using plenty of water. Otherwise, it can
damage the skin or clothes.
Do not pour more than 1 mL of detergent (CELLCLEAN) into the
chamber. It may overflow, possibly causing electric shocks or leakage.
CAUTION: Take care not to have detergent (CELLCLEAN) adhere to the chamber
side wall.
(5) Close the transducer cover.
(6) Close the front cover of the main unit.
WARNING
CAUTION