MAINTENANCE AND SUPPLIES REPLACEMENT
Sysmex KX-21 Operator’s Manual -- Revised October 1998 4-15
(7) Press the start switch. The transducer cleaning is executed.
*
Clean Transducer
*
Please wait.
Figure 4-5-9: Screen during Transducer Cleaning
(8) When the transducer cleaning is completed, the auto rinse and background check are
executed, then the system turns to the Ready status.
(9) Check to see that no background error occurs. Should background error occur,
execute the auto rinse.
NOTE: • When this operation is completed, the transducer cycle counter is
automatically reset. For the confirmation of transducer cycle number,
refer to
Chapter 4, Section 9: CONFIRMATION OF CYCLE
NUMBER
.