MAINTENANCE AND SUPPLIES REPLACEMENT
4-16 Sysmex KX-21 Operator’s Manual -- Revised October 1998
6. EVERY 3-MONTH MAINTENANCE AND PROCEDURE
6.1 Clean Sample Rotor Valve (SRV)
When the power is turned on, the message is displayed once every three months or for every
7500 samples to instruct the scheduled maintenance.
*
Scheduled Maint.
*
1:Exec.Clean
Clean the SRV.
Cycle No. after cleaning. 7567
Date last serviced 31/12/97
Reset the cycle counter for SRV,
and you can turn OFF the power.
[1]
[3]
Continue to start up. SRV cycle
counter will not be reset.
3:Cancel
Figure 4-6-1: Scheduled Maintenance Message Screen
When this message appears, press [1] key to select "1: Exec. Clean," and reset the cycle
counter. Then clean the sample rotor valve (SRV) by the following procedure:
When cleaning the sample rotor valve (SRV), always wear rubber
gloves. After completion of operation, wash hands with disinfectant. If
your hands are contaminated by blood, etc., infection of bacteria or the
like can occur.
Because CELLCLEAN is a strong alkaline detergent, take care not to
have it adhere to the skin or clothes. If the skin or clothes should come
in touch with it, flush it away using plenty of water. Otherwise, it can
damage the skin or clothes.
NOTE: When you press [3] key and select "3: Cancel," the start-up process
continues without cleaning the sample rotor valve (SRV). Until the
sample rotor valve (SRV) is cleaned, the above message is displayed
at start-up.
When you want to clean before this message appears, or when you
selected "3: Cancel" on the Scheduled Maint. screen and executed
cleaning when the power was off, reset the cycle counter, making
reference to
Chapter 4, Section 6.2: Reset SRV Cycle Counter
.