MAINTENANCE AND SUPPLIES REPLACEMENT
Sysmex KX-21 Operator’s Manual -- Revised October 1998 4-25
NOTE: • After Auto Rinse, background check is executed. When the background
value of any parameter exceeds tolerance, background check is
repeated a maximum of twice. In case background value still exceeds
tolerance, the message "Background Error" appears while the alarm
starts sounding. Press [HELP] key. This stops the alarm and the
Action Message screen appears. Follow the Action Message screen
and take appropriate action. (Refer to
Chapter 7:
TROUBLESHOOTING
).
• Background tolerance
WBC 0.3 [×10
3
/µL] or less
RBC 0.02 [×10
6
/µL] or less
HGB 0.1 [g/dL] or less
PLT 10 [×10
3
/µL] or less
Table 4-7-1: Background tolerance