MAINTENANCE AND SUPPLIES REPLACEMENT
4-26 Sysmex KX-21 Operator’s Manual -- Revised October 1998
7.2 Clean Rinse Cup
When the rinse cup has any blood on it or is clogged, clean it by the following procedure:
When cleaning the rinse cup, always wear rubber gloves. After
completion of operation, wash hands with disinfectant. If your hands
are contaminated by blood, etc., infection of bacteria or the like can
occur.
(1) Turn off the power of the main unit and wait approximately 30 seconds.
(2) Open the front cover of the main unit.
(3) Gently push down the rinse cup using both hands. Make sure the rinse cup is
removed completely from the sample probe.
Sample Probe
Rinse Cup
Figure 4-7-3: Pushing Down the Rinse Cup
(4) Remove the rinse cup in the order of 1, 2, and 3 as shown below.
Rinse Cup
3
2
1
Figure 4-7-4: Removing the Rinse Cup
WARNING