QUALITY CONTROL
Sysmex KX-21 Operator’s Manual -- Revised October 1998 5-23
CAUTION: • Print of anal
sis result is valid until the data is set. If the data is set b
pressin
1
ke
or
2
ke
, anal
sis result cannot be printed.
• This print format cannot be chan
ed since it is the special format for QC
anal
sis.
(9) Set the data using [1] key or [2] key.
[1: OK]: Sets QC data, and the plot is entered to the QC Chart. This QC data is kept
in the stored data.
[2: NG]: Cancels the analysis result, and the system remains to be ready for another
analysis.
CAUTION: • When an error occurs durin
QC anal
sis, error messa
e is displa
ed
in the s
stem status area.
Anal
sis data in which error occurred is masked, which cannot be
accepted as QC data. When
ou press
1
ke
to select "1: OK" in
response to the data confirmation messa
e, the alarm sounds and "1:
OK" is unacceptable.
Press
HELP
ke
and perform the error recover
process. For the error
recover
process, refer to
Chapter 7: TROUBLESHOOTING
.
(10) The set data is printed.
No. QC01
Date 31/12/97 10:02
Mode QC
WBC 7.3 ×10
3
/µL
RBC 4.52 ×10
6
/µL
HGB 14.8 g/dL
HCT 36.2 %
MCV 80.1 fL
MCH 32.7 Pg
MCHC 40.9 g/dL
PLT 206 ×10
3
/µL
LYM% 27.2 %
MXD% 13.4 %
NEUT% 59.4 %
LYM# 2.0 ×10
3
/µL
MXD# 1.0 ×10
3
/µL
NEUT# 4.3 ×10
3
/µL
W-SMV 56.3 fL
W-LMV 172.4 fL
RDW-CV 11.4 %
RDW-SD 27.4 fL
PDW 8.3 fL
MPV 9.1 fL
P-LCR 14.8 %
Figure 5-2-35: Example of Printing
Sysmex KX-21 Operator’s Manual -- Revised January 2000