CALIBRATION
6-2 Sysmex KX-21 Operator’s Manual -- Revised October 1998
1.4 Calibration Flow Chart
HGB/HCT calibration start
Completion of HGB/HCT calibration
Analyze samples by the reference method to determine
reference value (target value).
Automatic calibration: 5 samples
Manual calibration: Over 5 samples
Analyze in whole blood
mode.
Calculate calibration value.
Select manual calibration
program.
Enter calibration value.
Update calibration value.
Select automatic calibration
program.
Enter target value.
Analyze in whole blood
mode.
Update calibration value.
Select calibration parameter.
1: HGB
2: HGB and HCT
3: HCT
Page 6-1
Page 6-11
Page 6-12
Page 6-10
Page 6-9
Page 6-9
Page 6-3
Page 6-4
Page 6-5
Page 6-8
Page 6-6
Automatic
calibration
Manual
calibration