CALIBRATION
Sysmex KX-21 Operator’s Manual -- Revised October 1998 6-3
2. AUTOMATIC CALIBRATION
In automatic calibration, five or more fresh, normal blood samples are used for automatic
calibration of HGB and HCT values.
2.1 Executing Automatic Calibration Program
(1) Press [SELECT] key in the Ready status. The Select Menu screen appears.
(2) Using [ ] or [ ] key, move the cursor to select "3: Calibration."
(3) Press [ENTER] key. The Calibration Menu screen appears.
*
Calibration
*
1:Auto Cal.
2:Manual Cal.
Figure 6-2-1: Calibration Menu Screen
(4) Using [
] or [ ] key, move the cursor to select "1: Auto Cal."
(5) Press [ENTER] key. The Automatic Calibration Menu screen appears.
*
Auto Cal.
*
1:HGB
2:HGB/HCT
3:HCT
Figure 6-2-2: Automatic Calibration Menu Screen
On the Automatic Calibration Menu screen, an automatic calibration parameter can be selected
from among "1: HGB," "2: HGB/HCT," and "3: HCT."
NOTE: • Each menu can be selected directly by entering a numeral from the
numeric keys.