CALIBRATION
Sysmex KX-21 Operator’s Manual -- Revised October 1998 6-5
Target
HCT
Current Cal.(%)100.0
New Cal.(%)
1
2
3
4
5
Data
HCT
Compensation
HCT
*
Auto Cal.
*
Press [SELECT] to Set.
Figure 6-2-5: HCT Automatic Calibration Screen
2. Entering Target Values
(1) Using the numeric keys, enter the HGB or HCT reference values obtained by the
reference method in the target column.
NOTE: • Press [C] key to delete one letter at a time.
(2) Press [ENTER] key to fix the entered value and move the reverse display to the next
target column.
Target
HGB
12.5
HCT HCT HCT
Current Cal.(%)100.0 100.0
New Cal.(%)
1
2
3
4
5
Data
HGB
Compensation
HGB
*
Auto Cal.
*
Press [SELECT] to Set.
Figure 6-2-6: Entering Target Values
NOTE: • After entering the 5th target value, pressing [ENTER] key does not
move the reverse display to the next target column.
• An entered value can be set also by pressing [
] or [ ] key after
entering the target value.