CALIBRATION
6-10 Sysmex KX-21 Operator’s Manual -- Revised October 1998
3.2 Manual Calibration Procedure
(1) Press [SELECT] key in the Ready status. The Select Menu screen appears.
(2) Using [ ] or [ ] key, move the cursor to select "3: Calibration."
(3) Press [ENTER] key. The Calibration Menu screen appears.
*
Calibration
*
1:Auto Cal.
2:Manual Cal.
Figure 6-3-1: Calibration Menu Screen
(4) Using [ ] or [ ] key, move the cursor to select "2: Manual Cal."
(5) Press [ENTER] key. The Manual Calibration screen appears.
HGB
HCT
100.0
100.0
*
Manual Cal.
*
Press [SELECT] to exit.
Figure 6-3-2: Manual Calibration Screen