EasyManua.ls Logo

SYSMEX KX-21 - Page 189

SYSMEX KX-21
306 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TROUBLESHOOTING
Sysmex KX-21 Operator’s Manual -- Revised October 1998 7-15
3.4 Transducer Errors
Error Message Meaning of Message
"WBC Aperture Clog"
"RBC Aperture Clog"
Clog in the aperture
Parameter: WBC RBC
1 Execute clog removal sequence.
If clog persists,
press[SELECT],then
[Maintenance]->[Clean Transducer]
or dab with brush after [SELECT],
[Maintenance]->[Drain TD Chamber]
ERR CODE:XXXXX.X.X
1:Clog Remove
*
Help
*
3:Cancel
The monitor value for aperture clog is out of the
prescribed range.
Cause 1) Transducer aperture clogging
Action 1) Remove Aperture Clogging.
Press [1] key to execute automatic clog removal sequence.
2) Transducer Rinse Sequence
Pour 1 mL each of CELLCLEAN in the transducer chamber to
perform auto rinse sequence.
Refer to
Chapter 4, Section 5.2: Clean Transducer
.
3) Clean the Transducer with Brush
Clean the aperture with brush.
Refer to
Chapter 4, Section 7.3: Clean WBC/RBC Transducer
Aperture.
Analysis after error
occurrence
Ready for analysis (although the next sample could be affected)
NOTE: Press [3] key to return to the Analysis screen. Although the system is
ready for analysis, the error may occur in the next analysis.

Related product manuals