TROUBLESHOOTING
7-16 Sysmex KX-21 Operator’s Manual -- Revised October 1998
3.5 Temperature Errors
Error Message Meaning of Message
"Room Temp. High"
"Room Temp. Low"
ERR CODE:XXXXX.X.X
*
Help
*
Room Temp.Error
Analysis data has lower reliability.
Check the room temperature.
41.5˚C [10.0 - 40.0]
3:Return
Temperature in the unit (around the transducer)
is out of the monitor range.
Monitor range: 10.0 - 40.0°C
Cause 1) Temperature in the transducer is high (or low.)
Action 1) Check the Ambient Temperature
Check to see that the ambient temperature is between 15°C -
30°C.
Analysis after error
occurrence
Ready for analysis (although the next sample could be affected)
NOTE: Press [3] key to return to the Analysis screen. Although the system is
ready for analysis, the error may occur in the next analysis.